QUART DVTap

DIN 6868-161 CBCT Image Quality Test Phantom

Developed over a period of 2 years in a project involving major dental manufacturers, and released by QUART already in 2007, the DVTap has become a national and international standard solution for CBCT - based on our company's innovative approach.

The QUART DVTap phantom is designed to be used as a universal tool for 3D imaging equipment including CT applications. The phantom fully complies with DIN 6868-161 for acceptance tests in DVT/CBCT.

In conjunction with a specially developed software (QUART DVTtec), quick and comprehensive CBCT IQ tests can be performed.

Only one exposure is necessary to create a 3D data set containing all required parameters to evaluate CBCT image quality. Automated evaluation is performed with the unique QUART DVTtec software.

The phantom can be used for field sizes from 4x4cm to large fields-of-view (FOV). A universal holder or customised phantom holders are available for easy and reproducible positioning.

Technical specifications

  • SPATIAL RESOLUTION: Line spread function
  • RESOLUTION: Z-Resolution
  • STANDARD TEST OBJECTS: PMMA / Air / PVC
  • MATERIAL EQUIVALENTS: Free Air / Soft tissue / Bone
  • POSTIONING TOOLS: Linear (top side) / Selective markers
  • SIZE: Ø 16 cm, height: 15 cm
  • SCATTER RADIATION MODULES: 1x 6 cm / 1x 5 cm
Parameters
  • Nyquist Frequency (NF)
  • Contrast-to-Noise Ratio (CNR)
  • Homogeneity / Image Uniformity
  • Z-Resolution
  • Modulation Transfer Function (MTF)
  • Artefacts, Image Flaws
  • Figure of Merit / Acceptance Indicator

EFOMP working group, Quality control in cone-beam computed tomography (CBCT) EFOMP-ESTRO-IAEA protocol (summary report), Physica Medica, Volume 39, July 2017, Pages 67-72.