didoEASY MR / didoEASY+ MR / didoEASY++ MR

X-Ray Multi-Meter Series for Service and QA in R&F, Dental and Mammography
QUART didoEASY MR
Art. 11117
Basic Configuration
QUART didoEASY+ MR
Art. 11117+
Added kV Measurement
QUART didoEASY++ MR
Art. 11117++
Added kV and Direct-HVL Measurement

 

The QUART didoEASY meters are designed for users who emphasise high precision in dosimetric applications but do not require the performance of a full-range dosemeter package.

In basic configuration QUART didoEASY meters can be used to measure parameters which are essential for service and quality assurance operations at x-ray equipment such as dose, dose rate and time. Of course, as with all QUART meters – with maximum precision.

- kVp and PPV measurement are available in the QUART didoEASY+ MR version.

- Direct-HVL and kVp / PPV measurement are additionally available in the QUART didoEASY++ MR version.

Since the meters do not require any pre-setting procedure, measurement results are acquired very quickly:

1. Position the detector and switch on the didoEASY meter...

2. Set the x-ray equipment to the desired parameters...

3. Expose... - and simply read all measurement information from the meter‘s display.

The R detector automatically compensates all major medical x-ray beam qualities (RQR and RQA series (IEC 61267:2005)). No further corrections or compensations are required.

The M detector for mammography automatically compensates for Mo/Mo beam quality. Corrections factors are provided for further radiation qualities.

Like the QUART dido2000 series meters, the detector of the QUART didoEASY meter series also measures the integrated dose-width product (DWP) at dental panoramic equipment.* A feature that provides real added value for QUART users.

Technical specifications

General Features
  • Exposures: one to acquire all parameters
  • Mode: auto-start, auto-stop & auto-reset
  • Power supply: Rechargeable Li Ion accu
  • Battery time: approx. 120 hrs. until recharge required
  • Base size: 17x7x4.5cm (LxWxH)
  • System weight: 200 g
  • Detector size: 6.0x1.8x0.5cm (LxWxH)
  • Detector weight: negligible (!)

 

Specifications

 

didoEASY MR    basic configuration

  • Dose: 0.99 µGy – 999 mGy (Uncertainty: ±5% @ 25-40 kV (M) @ 50-150 kV (R) / Resolution: 0.01 µGy)
  • DLP*: 0.99 µGycm – 999 mGycm (Uncertainty: ±5% @ 50-120 kV / Resolution: 0.01 µGycm)
  • Dose Rate: 0.5 µGy/s – 999 mGy/s (Uncertainty: ±5% @ 50-150 kV / Resolution: 0.1 µGy/s)
  • Time: 1.0 ms - 296 s (Uncertainty: ±1.4 ms / Resolution: 0.1 ms)

 

didoEASY+  MR   - features added kV measurement 

  • Dose: 0.99 µGy – 999 mGy (Uncertainty: ±5% @ 25-40 kV (M) @ 50-150 kV (R) / Resolution: 0.01 µGy)
  • DLP*: 0.99 µGycm – 999 mGycm (Uncertainty: ±5% @ 50-120 kV / Resolution: 0.01 µGycm)
  • Dose Rate: 0.5 µGy/s – 999 mGy/s (Uncertainty: ±5% @ 50-150 kV / Resolution: 0.1 µGy/s)
  • Time: 1.0 ms - 296 s (Uncertainty: ±1.4 ms / Resolution: 0.1 ms)
  • Pulses: 1 - 9999 (Uncertainty: ± 1 pulse)
  • Pulse Rate: 0.1 - 250 pulses/s (Uncertainty: ± 0.1 pulses/s)
  • kVp / PPV
  • 25 – 35 kV (applies to M detectors)
  • 50 – 130 kV (applies to R detectors)
  • (Uncertainty: ±5% at calibrated reference points / Resolution: 0.1 kV); Exposure conditions: Mammo: Mo/Mo; HVL 0.2 – 0.45 mm Al eq., correction factors provided for further radiation qualities / R&F: HVL 1.5 – 10.5 mm Al eq.

 

didoEASY++  MR   - features added kV and direct-HVL measurement

  • Dose: 0.99 µGy – 999 mGy (Uncertainty: ±5% @ 25-40 kV (M) @ 50-150 kV (R) / Resolution: 0.01 µGy)
  • DLP*: 0.99 µGycm – 999 mGycm (Uncertainty: ±5% @ 50-120 kV / Resolution: 0.01 µGycm)
  • Dose Rate: 0.5 µGy/s – 999 mGy/s (Uncertainty: ±5% @ 50-150 kV / Resolution: 0.1 µGy/s)
  • Time: 1.0 ms - 296 s (Uncertainty: ±1.4 ms / Resolution: 0.1 ms)
  • Pulses: 1 - 9999 (Uncertainty: ± 1 pulse)
  • Pulse Rate: 0.1 - 250 pulses/s (Uncertainty: ± 0.1 pulses/s)
  • kVp / PPV
  • 25 – 35 kV (applies to M detectors)
  • 40 – 125 kV (applies to R detectors)
  • (Uncertainty: ±5% at calibrated reference points / Resolution: 0.1 kV); Exposure conditions: Mammo: Mo/Mo; HVL 0.2 – 0.45 mm Al eq., correction factors provided for further radiation qualities / R&F: HVL 1.5 – 10.5 mm Al eq.
  • Direct-HVL: (R) 1.1 – 13.3 mm Al (HVL output, Uncertainty: ±10% / Resolution: 0.1 mm Al); Internal filtration 1.5 – 50 mm Al; 40-125 kV / (M) 0.2 – 0.45 mm Al (HVL output, Uncertainty: ±10% / Resolution: 0.1 mm Al); Internal filtration Mo/Mo; 25-35 kV

* S A Mitchell and C J Martin, Comparison of ionisation chamber and semiconductor detector devices for measurement of the dose–width product for panoramic dental units, J. Radiol. Prot. 33 321 (2013).